RecallRadar

FDA Device recall Z-1709-2022

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518

On 2022-09-21 the FDA classified a Class II recall of Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518 by Cook, citing product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1709-2022
ClassificationClass II
Statusongoing
Initiated2022-07-19
Published2022-09-21
Last updated2026-09-13 11:56 UTC
CompanyCook
DistributionUS

Product

Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.

Reason

Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.

Identifiers

code_infoUDI-DI: (01)00827002075184 Lot Numbers: 14063361, 14087089, 14087092, 14117033, 14120926, 14120928, 14151216, 14151218, 14153159, 14158614, 14159695, 14159696

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1709-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.