FDA Device recall Z-1708-2022
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516
- FDA Device
- Class II
- ongoing
- Published 2022-09-21
On 2022-09-21 the FDA classified a Class II recall of Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516 by Cook, citing product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1708-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-07-19 |
| Published | 2022-09-21 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Cook |
| Distribution | US |
Product
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN):RLPC-35-180 ORDER NUMBER (GPN) G07516. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Reason
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Identifiers
| code_info | UDI-DI: (01)00827002075160- Lot Numbers: 14103953, 14103956, 14143728, 14145412, 14145419, 14153927, 14156907, 14156908, 14156914, 14160996 |
|---|