FDA Device recall Z-1702-2014

LifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by intravenous (I.V.) administr…

On 2014-06-11 the FDA classified a Class II recall of LifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by intravenous (I.V.) administr… by Hospira, citing there is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1702-2014
ClassificationClass II
Statusterminated
Initiated2014-01-24
Published2014-06-11
Last updated2026-09-13 11:53 UTC
CompanyHospira
DistributionUS

Product

LifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by intravenous (I.V.) administration.

Reason

There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion

Identifiers

code_infoList Number: 02507; Serial Numbers: All.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.