FDA Device recall Z-1701-2018
MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014…
- FDA Device
- Class I
- terminated
- Published 2018-05-23
On 2018-05-23 the FDA classified a Class I recall of MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014… by Micro Therapeutics, citing there is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1701-2018 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2018-02-26 |
| Published | 2018-05-23 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Micro Therapeutics |
| Distribution | n/a |
Product
MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.
Reason
There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.
Identifiers
| code_info | All lot numbers: Product Number: UDI (a) REF 300010: 00843889031334, 00843889031341, 00847536020207 (b) REF 300011:…All lot numbers: Product Number: UDI (a) REF 300010: 00843889031334, 00843889031341, 00847536020207 (b) REF 300011: 00843889031365, 00843889031372, 00847536020214 (c) REF 300012: 00843889031396, 00843889031402, 00847536020221 (d) REF 300013: 00843889031426, 00843889031433, 00847536020238 (e) REF 300014: NOT DISTRIBUTED (f) REF 300015: 00847536020122, 00847536020160 (g) REF 300016: 00847536020139, 00847536020177 (h) REF 300017: 00847536020146, 00847536020184 (i) REF 300018: 00847536020153, 00847536020191 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1701-2018%22