RecallRadar

FDA Device recall Z-1699-2015

M/L Taper with Kinectiv¿ Technology. prosthesis, hip, semi-constrained (metal uncemented acetabular component) Product Usage: Usage: Total…

On 2015-06-17 the FDA classified a Class I recall of M/L Taper with Kinectiv¿ Technology. prosthesis, hip, semi-constrained (metal uncemented acetabular component) Product Usage: Usage: Total… by Zimmer, citing zimmer is initiating a voluntary recall of 64 lots (752 implants total) of M/L Taper with Kinectiv¿ femoral stems and modular necks due higher than allowed cytotoxicity levels fou….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1699-2015
ClassificationClass I
Statusterminated
Initiated2015-05-18
Published2015-06-17
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

M/L Taper with Kinectiv¿ Technology. prosthesis, hip, semi-constrained (metal uncemented acetabular component) Product Usage: Usage: Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures.

Reason

Zimmer is initiating a voluntary recall of 64 lots (752 implants total) of M/L Taper with Kinectiv¿ femoral stems and modular necks due higher than allowed cytotoxicity levels found with the product. Reasonable probability of adverse biological response and subsequent revision

Identifiers

code_info
Product: 00771300500 Lots: 63006851 63006852 63024180 Product: 00771300600 Lots: 63024183 63024184 6302418…Product: 00771300500 Lots: 63006851 63006852 63024180 Product: 00771300600 Lots: 63024183 63024184 63024186 63024187 63024188 63024189 Product: 00771300700 Lots: 63024193 63024195 63024196 63024197 63024198 63024199 63024201 63024202 63024203 63024204 63024205 63024206 Product: 00771300900 Lots: 62927082 62927083 63024210 63024211 63024213 63024214 63024215 63024216 63024217 63024218 63024219 63024220 63024221 Product: 00771301000 Lots: 62938997 63024226 63024227 63024228 63024229 63024230 Product: 00771301100 Lots: 62885040 62905574 62998426 63024234 63024235 63024236 63024237 63024238 63024239 63024240 63024241 Product: 00771301200 Lots: 62927123 63024256 63024257 63024258 63024259 63024261 63024262 63024263 Product: 00771301300 Lots: 62885058 62939008 63024245 Product: 00784801400 Lot: 62924878 Product: 65771301100 Lot: 62939041

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1699-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.