RecallRadar

FDA Device recall Z-1697-2025

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/A Product Description:…

On 2025-05-21 the FDA classified a Class I recall of Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/A Product Description:… by Vyaire Medical, citing adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hy….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1697-2025
ClassificationClass I
Statusongoing
Initiated2025-04-10
Published2025-05-21
Last updated2026-09-13 11:57 UTC
CompanyVyaire Medical
DistributionUS

Product

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit dual-limb, dual-heat, high-flow circuit (>4 L/min) Component: No. a conduit for respiratory gas between the patient and a ventilator.

Reason

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

Identifiers

code_info
Lot Code: Lot/Serial Number(s): 0004240347 0004240348 0004252021 0004253194 0004253470 0004255176 0004260100…Lot Code: Lot/Serial Number(s): 0004240347 0004240348 0004252021 0004253194 0004253470 0004255176 0004260100 0004262183 0004262987 0004263371 0004292077 0004300092 0004301668 UDI-DI AirLife Label Each: 10889483595862 Case: 30889483595866 Vyaire Label Each: 10190752145139 Case: 50190752145137

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1697-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.