FDA Device recall Z-1692-2021

Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or lo…

On 2021-06-09 the FDA classified a Class II recall of Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or lo… by Draeger Medical, citing three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1692-2021
ClassificationClass II
Statusterminated
Initiated2021-04-16
Published2021-06-09
Last updated2026-09-13 11:56 UTC
CompanyDraeger Medical
DistributionUS

Product

Critical Care Ventilator, Catalog Number(s): 8422200: Babylog VN600 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

Reason

Three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning

Identifiers

code_infoAll lots/serial numbers manufactured from 1/1/2019-3/31201

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1692-2021%22