FDA Device recall Z-1692-2017

UOC Femoral Driver, Product number PE 93045101, orthopedic manual surgical instrument/accessory for use during surgery to implant femoral c…

On 2017-04-05 the FDA classified a Class II recall of UOC Femoral Driver, Product number PE 93045101, orthopedic manual surgical instrument/accessory for use during surgery to implant femoral c… by United Orthopedic, citing the locking lever pin component of the femoral driver may become deformed and break, resulting in malfunction during surgery and causing the device to become inoperable.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1692-2017
ClassificationClass II
Statusterminated
Initiated2017-02-10
Published2017-04-05
Last updated2026-09-13 11:54 UTC
CompanyUnited Orthopedic
DistributionCA

Product

UOC Femoral Driver, Product number PE 93045101, orthopedic manual surgical instrument/accessory for use during surgery to implant femoral components and femoral trial, and to remove femoral trial.

Reason

The locking lever pin component of the femoral driver may become deformed and break, resulting in malfunction during surgery and causing the device to become inoperable.

Identifiers

code_infoLot numbers T16A178A (17 units); T16A178B (1 unit), T16A178B1 (3 units); T16A178C (12 units)