SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107

On 2015-06-03 the FDA classified a Class II recall of SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107 by Biosense Webster, citing biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an RMT procedure the SMARTABLATE RF Gener….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1691-2015
ClassificationClass II
Statusterminated
Initiated2015-03-19
Published2015-06-03
Last updated2026-09-13 11:53 UTC
CompanyBiosense Webster
DistributionUS

Product

SMARTABLATE RF Generator System, Catalog No. M490006, M490007, M4900106, and M4900107. Product Usage: The use of the SMART ABLATE Generator and all accessories is indicated in combination with compatible therapeutic catheters for use in conventional intracardiac RF ablation procedures. The SMARTABLATE Foot Pedal is designed to initiate or discontinue operation of the SMARTABLATE System. The Foot Pedal can be connected to the SMARTABLATE RF Generator, SMARTABLATE Irrigation Pump; or SMARTABLATE Remote Control.

Reason

Biosense Webster is recalling the SMARTABLATE Generator System because Biosense Webster has received two complaints reporting that during an RMT procedure the SMARTABLATE RF Generator inadvertently delivered RF energy.

Identifiers

code_info
SMARTABLATE System Serial No. G4C-0296, G4C-0297, G4C-0272, G4C-0262, G4C-0232, G4C-0090. System Serial No. G4C-0076…SMARTABLATE System Serial No. G4C-0296, G4C-0297, G4C-0272, G4C-0262, G4C-0232, G4C-0090. System Serial No. G4C-0076, G4C-0195, G4C-0064, G4C-0047, G4C-0159, G4C-0187, G4C-0146, G4C-0140, G4C-0037, G4C-0131, G4C-0031, G0054.