RecallRadar

FDA Device recall Z-1690-2024

Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterrupt…

On 2024-05-01 the FDA classified a Class II recall of Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterrupt… by Philips North America, citing possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down d….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1690-2024
ClassificationClass II
Statusongoing
Initiated2024-03-22
Published2024-05-01
Last updated2026-09-13 11:57 UTC
CompanyPhilips North America
DistributionUS

Product

Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model: SCL500RMI1U

Reason

Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition

Identifiers

code_info
1. SCL500RM1U (only serial number date codes prior to 2237*) 2. SCL500RMI1U (only serial number date codes prior to 2…1. SCL500RM1U (only serial number date codes prior to 2237*) 2. SCL500RMI1U (only serial number date codes prior to 2241*) Example unit serial number is 5S"2104"08943, meaning it was manufactured during week 04 of the year 2021. The serial number format is 5SYYWW123456.