RecallRadar

FDA Device recall Z-1688-2025

IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170

On 2025-05-07 the FDA classified a Class II recall of IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170 by Boston Scientific, citing boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1688-2025
ClassificationClass II
Statusongoing
Initiated2025-03-24
Published2025-05-07
Last updated2026-09-13 11:57 UTC
CompanyBoston Scientific
DistributionUS

Product

IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.

Reason

Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.

Identifiers

code_info
GTIN 00191506032692, Batch Numbers: 33744276, 33744277, 33744278, 33744600, 33931638, 33931639, 33931900, 33980480, 33…GTIN 00191506032692, Batch Numbers: 33744276, 33744277, 33744278, 33744600, 33931638, 33931639, 33931900, 33980480, 33980481, 33980482, 33989380, 33989381, 33989382, 34063066, 34063067, 34063068, 34063069, 34063070, 34084080, 34084081, 34084084, 34094111, 34094112, 34110333, 34110334, 34110335, 34110336, 34110337, 34168777, 34168778, 34204925, 34204926, 34204927, 34295709.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1688-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.