RecallRadar

FDA Device recall Z-1687-2019

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

On 2019-06-05 the FDA classified a Class II recall of Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only by Abbott Vascular, citing reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implanta….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1687-2019
ClassificationClass II
Statusterminated
Initiated2019-05-01
Published2019-06-05
Last updated2026-09-13 11:55 UTC
CompanyAbbott Vascular
DistributionAL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV

Product

Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Reason

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.

Identifiers

code_infoAll Serial Numbers.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.