Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)

On 2022-09-14 the FDA classified a Class II recall of Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F) by Saranas, citing bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1684-2022
ClassificationClass II
Statusongoing
Initiated2022-08-10
Published2022-09-14
Last updated2026-09-13 11:56 UTC
CompanySaranas
DistributionUS

Product

Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)

Reason

Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.

Identifiers

code_info
UDI/REF/Lot: 00866532000402/FG001-01/510090S, 511110A, 510120F, 102422, 102445, 102531, 102671, 102685, 102731; 0086653…UDI/REF/Lot: 00866532000402/FG001-01/510090S, 511110A, 510120F, 102422, 102445, 102531, 102671, 102685, 102731; 00866532000419/FG001-02/515070A, 528090C, 510120G, 511110B, 102567, 102610, 102611