FDA Device recall Z-1682-2021

Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide opt…

On 2021-06-02 the FDA classified a Class II recall of Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide opt… by Pentax, citing IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1682-2021
ClassificationClass II
Statusterminated
Initiated2021-04-21
Published2021-06-02
Last updated2026-09-13 11:56 UTC
CompanyPentax
DistributionUS

Product

Pentax Medical Video Duodenoscope - Product Usage: intended to be used with endoscopic devices and other ancillary equipment to provide optical visualization of (via a video monitor), and therapeutic access to, Biliary Tract via the Upper Gastrointestinal Tract, Model Number: ED34-i10T2.

Reason

IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration

Identifiers

code_infoAll units