FDA Device recall Z-1678-2014

IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001

On 2014-06-11 the FDA classified a Class II recall of IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001 by Philips Medical Systems Cleveland, citing during SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. I….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1678-2014
ClassificationClass II
Statusterminated
Initiated2014-03-11
Published2014-06-11
Last updated2026-09-13 11:53 UTC
CompanyPhilips Medical Systems Cleveland
DistributionUS

Product

IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT, Model number 881035), Philips Medical Systems. A medical software system for diagnostic imaging data viewing.

Reason

During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.

Identifiers

code_info
ISP DX/HX/EX; Model number 881001, Serial numbers 35033, 85298, 80861, 81017, 81021, 81191, 82085, 82124, 82144, 82148,…ISP DX/HX/EX; Model number 881001, Serial numbers 35033, 85298, 80861, 81017, 81021, 81191, 82085, 82124, 82144, 82148, 82242, 85029, 85030, 85033, 85164, 85166, 85183, 85247, 85302, 85324, 85337, 85353, 85357, 85359, 85397, 85401, 85402, 87004, 87014, 87158, 87176, 87177, 87182, 87200, 92003, 95043, 95044, 95045, 95046, 95047, 95048, 95049, 95050, 95051, 95052, 95054, 850001, 870030, 870071, 870073, 870076, 920007, 920008, 950006, 950023, 950031, 950039, 950041, 950045, 950052, 950071, 950084, 950085, 950087, 950106, 950119, 950120, 950121, 950129, 950137, 950143 and 950146. ISP IX; Model number 881030, Serial numbers 40024, 40021, 40031, 40036, 40044, 40061, 40082, 40088, 40096, 40145, 40146, 40148, 40150, 40155, 40158, 40160, 40169 and 600003. ISP LX SPECT; Model number 881035, Serial numbers 40001, 40003, 40033, 40084, 40087, 40093, 40133, 40134 and 40141.