FDA Device recall Z-1678-2014
IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001
- FDA Device
- Class II
- terminated
- Published 2014-06-11
On 2014-06-11 the FDA classified a Class II recall of IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001 by Philips Medical Systems Cleveland, citing during SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. I….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1678-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-03-11 |
| Published | 2014-06-11 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Philips Medical Systems Cleveland |
| Distribution | US |
Product
IntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT, Model number 881035), Philips Medical Systems. A medical software system for diagnostic imaging data viewing.
Reason
During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.
Identifiers
| code_info | ISP DX/HX/EX; Model number 881001, Serial numbers 35033, 85298, 80861, 81017, 81021, 81191, 82085, 82124, 82144, 82148,…ISP DX/HX/EX; Model number 881001, Serial numbers 35033, 85298, 80861, 81017, 81021, 81191, 82085, 82124, 82144, 82148, 82242, 85029, 85030, 85033, 85164, 85166, 85183, 85247, 85302, 85324, 85337, 85353, 85357, 85359, 85397, 85401, 85402, 87004, 87014, 87158, 87176, 87177, 87182, 87200, 92003, 95043, 95044, 95045, 95046, 95047, 95048, 95049, 95050, 95051, 95052, 95054, 850001, 870030, 870071, 870073, 870076, 920007, 920008, 950006, 950023, 950031, 950039, 950041, 950045, 950052, 950071, 950084, 950085, 950087, 950106, 950119, 950120, 950121, 950129, 950137, 950143 and 950146. ISP IX; Model number 881030, Serial numbers 40024, 40021, 40031, 40036, 40044, 40061, 40082, 40088, 40096, 40145, 40146, 40148, 40150, 40155, 40158, 40160, 40169 and 600003. ISP LX SPECT; Model number 881035, Serial numbers 40001, 40003, 40033, 40084, 40087, 40093, 40133, 40134 and 40141. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1678-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.