FDA Device recall Z-1675-2022

VITROS XT 3400 Chemistry System with Version 3.7.2 and below Product Code: 6844458

On 2022-09-14 the FDA classified a Class II recall of VITROS XT 3400 Chemistry System with Version 3.7.2 and below Product Code: 6844458 by Ortho Clinical Diagnostics, citing performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1675-2022
ClassificationClass II
Statusongoing
Initiated2022-07-19
Published2022-09-14
Last updated2026-09-13 11:56 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS XT 3400 Chemistry System with Version 3.7.2 and below Product Code: 6844458

Reason

Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.

Identifiers

code_infoUDI-DI: 10758750031986 Systems start with 3400XXXX where XXXX is unique for each analyzer

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.