RecallRadar

FDA Device recall Z-1675-2015

Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm

On 2015-06-03 the FDA classified a Class II recall of Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm by Mindray Ds Usa, citing software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously di….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1675-2015
ClassificationClass II
Statusterminated
Initiated2015-03-12
Published2015-06-03
Last updated2026-09-13 11:53 UTC
CompanyMindray Ds Usa
DistributionUS

Product

Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.

Reason

Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).

Identifiers

code_infoPart No: 0998-00-0708-01 Software version 8.9

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1675-2015%22