FDA Device recall Z-1675-2015
Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm
- FDA Device
- Class II
- terminated
- Published 2015-06-03
On 2015-06-03 the FDA classified a Class II recall of Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm by Mindray Ds Usa, citing software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously di….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1675-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-03-12 |
| Published | 2015-06-03 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Mindray Ds Usa |
| Distribution | US |
Product
Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
Reason
Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Identifiers
| code_info | Part No: 0998-00-0708-01 Software version 8.9 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1675-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.