FDA Device recall Z-1671-2020

ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizi…

On 2020-04-15 the FDA classified a Class II recall of ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizi… by Abbott Laboratories, citing mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1671-2020
ClassificationClass II
Statusterminated
Initiated2019-05-01
Published2020-04-15
Last updated2026-09-13 11:55 UTC
CompanyAbbott Laboratories
DistributionUS

Product

ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes.

Reason

Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.

Identifiers

code_infoLN 09D59-01, LN 09D59-02, LN 09D59-03

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1671-2020%22