RecallRadar

FDA Device recall Z-1670-2023

Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial)

On 2023-06-07 the FDA classified a Class II recall of Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial) by Smith And Nephew, citing recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a pote….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1670-2023
ClassificationClass II
Statusongoing
Initiated2023-05-02
Published2023-06-07
Last updated2026-09-13 11:56 UTC
CompanySmith And Nephew
DistributionUS

Product

Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); c) REF 1-10003-200 (SIZE 2-Right Medial); d) REF 1-10003-210 (SIZE 2-Left Medial); e) REF 1-10003-300 (SIZE 3- Right Medial); f) REF 1-10003-310 (SIZE 3- Left Medial); g) REF 1-10003-400 (SIZE 4- Right Medial); h) REF 1-10003-410 (SIZE 4- Left Medial); i) REF 1-10003-500 (SIZE 5- Right Medial); j) REF 1-10003-510 (SIZE 5- Left Medial); k) REF 1-10003-600 (SIZE 6- Right Medial); l) REF 1-10003-610 (SIZE 6- Left Medial); m) REF 1-10003-700 (SIZE 7- Right Medial); n) REF 1-10003-710 (SIZE 7- Left Medial); o) REF 1-10003-800, (SIZE 8 - Right Medial); p) REF 1-10003-810, (SIZE 8 - Left Medial); Unicondylar knee prothesis

Reason

Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

Identifiers

code_info
a) REF 1-10003-100, UDI/DI 00885556872925; b) REF 1-10003-110, UDI/DI 00885556872932; c) REF 1-10003-200, UDI/DI 00885…a) REF 1-10003-100, UDI/DI 00885556872925; b) REF 1-10003-110, UDI/DI 00885556872932; c) REF 1-10003-200, UDI/DI 00885556872949; d) REF 1-10003-210, UDI/DI 00885556872956; e) REF 1-10003-300, UDI/DI 00885556872963; f) REF 1-10003-310, UDI/DI 00885556872970; g) REF 1-10003-400, UDI/DI 00885556872987; h) REF 1-10003-410, UDI/DI 00885556872994; i) REF 1-10003-500, UDI/DI 00885556873007; j) REF 1-10003-510, UDI/DI 00885556873014; k) REF 1-10003-600, UDI/DI 00885556873021; l) REF 1-10003-610, UDI/DI 00885556873038; m) REF 1-10003-700, UDI/DI 00885556873045; n) REF 1-10003-710, UDI/DI 00885556873052; o) REF 1-10003-800, UDI/DI 00885556873069; p) REF 1-10003-810, UDI/DI 00885556873076; ALL BATCHES

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1670-2023%22