FDA Device recall Z-1670-2016

ADVIA Chemistry Triglyceride_2, concentrated

On 2016-05-25 the FDA classified a Class II recall of ADVIA Chemistry Triglyceride_2, concentrated by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1670-2016
ClassificationClass II
Statusterminated
Initiated2016-03-07
Published2016-05-25
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionCA, CO, CT, FL, KS, MI, MN, NC, NM, NY, OK, OR, TN, TX, UT, WA

Product

ADVIA Chemistry Triglyceride_2, concentrated; TRIG_c; Catalog # SMN # 10697575. For in vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma on ADVIA Chemistry systems.

Reason

Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932.

Identifiers

code_infoKit lots 348297 & 359932.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.