RecallRadar

FDA Device recall Z-1667-2019

Maquet Getinge- (1) Material: 701067350R01 BEQ-TOP 5208 ECC Small-Cardiac (2) material: 701067350R02 BEQ-TOP 5208 ECC Small-Cardiac

On 2019-06-05 the FDA classified a Class II recall of Maquet Getinge- (1) Material: 701067350R01 BEQ-TOP 5208 ECC Small-Cardiac (2) material: 701067350R02 BEQ-TOP 5208 ECC Small-Cardiac by Maquet Cardiovascular, citing custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1667-2019
ClassificationClass II
Statusterminated
Initiated2019-04-18
Published2019-06-05
Last updated2026-09-13 11:55 UTC
CompanyMaquet Cardiovascular
DistributionUS

Product

Maquet Getinge- (1) Material: 701067350R01 BEQ-TOP 5208 ECC Small-Cardiac (2) material: 701067350R02 BEQ-TOP 5208 ECC Small-Cardiac

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Identifiers

code_infoBatch Numbers: (1) 3000067384, 3000070970 (2) 3000074734, 3000081003

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.