FDA Device recall Z-1667-2018
Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRE Product Usage: The Baxter Sigma Spectrum Infusion Pump…
- FDA Device
- Class II
- terminated
- Published 2018-05-16
On 2018-05-16 the FDA classified a Class II recall of Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRE Product Usage: The Baxter Sigma Spectrum Infusion Pump… by Baxter Healthcare, citing service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitt….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1667-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-04-03 |
| Published | 2018-05-16 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Baxter Healthcare |
| Distribution | US |
Product
Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRE Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.
Reason
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.
Identifiers
| code_info | Serial number 758477 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1667-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.