FDA Device recall Z-1659-2018

Dimension Vista Creatine Kinase MB (MBI) assay

On 2018-05-16 the FDA classified a Class II recall of Dimension Vista Creatine Kinase MB (MBI) assay by Siemens Healthcare Diagnostics, citing various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely el….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1659-2018
ClassificationClass II
Statusterminated
Initiated2017-06-29
Published2018-05-16
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionDC, GA, PR

Product

Dimension Vista Creatine Kinase MB (MBI) assay; (Product Number: K3032/10464339) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)

Reason

Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.