ADVIA CHEMISTRY GLUCOSE OXIDASE ASSAY AND ADVIA Chemistry Glucose & Concentrated Glucose Oxidase Assay and ADVIA Chemistry Glucose

On 2016-05-18 the FDA classified a Class II recall of ADVIA CHEMISTRY GLUCOSE OXIDASE ASSAY AND ADVIA Chemistry Glucose & Concentrated Glucose Oxidase Assay and ADVIA Chemistry Glucose by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirm….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1655-2016
ClassificationClass II
Statusterminated
Initiated2016-03-17
Published2016-05-18
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA CHEMISTRY GLUCOSE OXIDASE ASSAY AND ADVIA Chemistry Glucose & Concentrated Glucose Oxidase Assay and ADVIA Chemistry Glucose.

Reason

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Identifiers

code_infoDevice Listing: Glucose Oxidase (GLUO) D11793, Catalog # 74020 & Concentrated Glucose Oxidase (GLUO_c) D117943, Catalog # 74721