FDA Device recall Z-1653-2019
Maquet Getinge- (1) Material: 701055490R01 BEQ-TOP 15801 ¿ inch ECC Pack (2) Material :701055490R02 BEQ-TOP 15801 1/4 in ECC Pack
- FDA Device
- Class II
- terminated
- Published 2019-06-05
On 2019-06-05 the FDA classified a Class II recall of Maquet Getinge- (1) Material: 701055490R01 BEQ-TOP 15801 ¿ inch ECC Pack (2) Material :701055490R02 BEQ-TOP 15801 1/4 in ECC Pack by Maquet Cardiovascular, citing custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1653-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-18 |
| Published | 2019-06-05 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Maquet Cardiovascular |
| Distribution | US |
Product
Maquet Getinge- (1) Material: 701055490R01 BEQ-TOP 15801 ¿ inch ECC Pack (2) Material :701055490R02 BEQ-TOP 15801 1/4 in ECC Pack
Reason
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Identifiers
| code_info | Batch Numbers: (1) 3000050886 (2) 3000066508, 3000081906 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1653-2019%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.