RecallRadar

FDA Device recall Z-1651-2016

Advia Chemistry Assays - Cholesterol_2 concentrated and Cholesterol, LDL Cholesterol

On 2016-05-18 the FDA classified a Class II recall of Advia Chemistry Assays - Cholesterol_2 concentrated and Cholesterol, LDL Cholesterol by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirm….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1651-2016
ClassificationClass II
Statusterminated
Initiated2016-03-17
Published2016-05-18
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Advia Chemistry Assays - Cholesterol_2 concentrated and Cholesterol, LDL Cholesterol

Reason

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Identifiers

code_infoCholesterol_2 (CHOL2): 74719 Cholesterol_concentrated (CHOL_c): 74718

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1651-2016%22