FDA Device recall Z-1650-2013

VITROS 3600 Immunodiagnostic System (product code 6802783) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qual…

On 2013-07-10 the FDA classified a Class II recall of VITROS 3600 Immunodiagnostic System (product code 6802783) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qual… by Ortho Clinical Diagnostics, citing there is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1650-2013
ClassificationClass II
Statusterminated
Initiated2013-05-06
Published2013-07-10
Last updated2026-09-13 11:53 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS 3600 Immunodiagnostic System (product code 6802783) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents

Reason

There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.

Identifiers

code_infoSerial numbers: J36000111 and J36000520 through J36000617

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.