FDA Device recall Z-1649-2013
VITROS 5600 Integrated System (product code 6802413) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitativ…
- FDA Device
- Class II
- terminated
- Published 2013-07-10
On 2013-07-10 the FDA classified a Class II recall of VITROS 5600 Integrated System (product code 6802413) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitativ… by Ortho Clinical Diagnostics, citing there is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1649-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-05-06 |
| Published | 2013-07-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS 5600 Integrated System (product code 6802413) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Reason
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Identifiers
| code_info | Serial numbers: J56000110, J56000118 and J56001095 through J56001507 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1649-2013%22