RecallRadar

FDA Device recall Z-1647-2018

ADVIA Alanine Aminotransferase (ALT) reagent chemistry assays

On 2018-05-16 the FDA classified a Class II recall of ADVIA Alanine Aminotransferase (ALT) reagent chemistry assays by Siemens Healthcare Diagnostics, citing various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely el….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1647-2018
ClassificationClass II
Statusterminated
Initiated2017-06-29
Published2018-05-16
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionDC, GA, PR

Product

ADVIA Alanine Aminotransferase (ALT) reagent chemistry assays; ALT intended to measure the activity of the enzyme alanine amino transferase (ALT) (also known as a serum glutamic pyruvic transaminase or SGPT) in serum and plasma. (Product Numbers: 10318168, 10309500), ALT P5P (Part Numbers: 10318168/10326245, 10309500/10315181), and ALT Concentrated (Part Numbers: 10283341, 10283341, 10283342); (Mfr: Randox Laboratories)

Reason

Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.

Identifiers

code_infoAll lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1647-2018%22