RecallRadar

FDA Device recall Z-1647-2014

Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CS…

On 2014-05-28 the FDA classified a Class II recall of Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CS… by Siemens Healthcare Diagnostics, citing A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1647-2014
ClassificationClass II
Statusterminated
Initiated2014-03-19
Published2014-05-28
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista¿ System. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.

Reason

A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C

Identifiers

code_info
CSA 13114BA, exp. 2014-04-24; CSA 13157BA, exp. 2014-06-06; CSA 13246BB, exp. 2014-09-03; CSA 13317BA, exp. 2014-11-13…CSA 13114BA, exp. 2014-04-24; CSA 13157BA, exp. 2014-06-06; CSA 13246BB, exp. 2014-09-03; CSA 13317BA, exp. 2014-11-13; CSA 14008BB, exp. 2015-01-08; CSA 14050BB, exp. 2015-02-19; CSAE 13157BC, exp. 2014-06-06; CSAE 13249BD, exp. 2014-09-06.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1647-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.