FDA Device recall Z-1647-2013
1) VITROS 5,1 FS Chemistry System (product code 6801375) and 2) VITROS 5,1 FS Chemistry System Refurbished (product code 6801890) Product U…
- FDA Device
- Class II
- terminated
- Published 2013-07-10
On 2013-07-10 the FDA classified a Class II recall of 1) VITROS 5,1 FS Chemistry System (product code 6801375) and 2) VITROS 5,1 FS Chemistry System Refurbished (product code 6801890) Product U… by Ortho Clinical Diagnostics, citing there is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1647-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-05-06 |
| Published | 2013-07-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
1) VITROS 5,1 FS Chemistry System (product code 6801375) and 2) VITROS 5,1 FS Chemistry System Refurbished (product code 6801890) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Reason
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Identifiers
| code_info | Serial numbers: 1) J34000240, J34000714, J34001562 and J34002163 through J34002245, 2) 34000193, 34000420, 34000478, 34…Serial numbers: 1) J34000240, J34000714, J34001562 and J34002163 through J34002245, 2) 34000193, 34000420, 34000478, 34000495, 34000497, 34000529, 34000596, 34000799, 34000805, 34000947, 34001243, 34001422, 34001504, 34001569, 34001589, 34001610, 34001633 |
|---|