FDA Device recall Z-1645-2018

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on…

On 2018-05-16 the FDA classified a Class II recall of IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on… by Deerfield Imaging, citing the DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1645-2018
ClassificationClass II
Statusterminated
Initiated2018-04-04
Published2018-05-16
Last updated2026-09-13 11:54 UTC
CompanyDeerfield Imaging
DistributionUS

Product

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designated to operative inside an RF shielded room to facilitate intraoperative and multi room use.

Reason

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

Identifiers

code_infoSerial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.