RecallRadar

FDA Device recall Z-1642-2026

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The P…

On 2026-04-01 the FDA classified a Class II recall of Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The P… by Olympus, citing olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Fo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1642-2026
ClassificationClass II
Statusongoing
Initiated2026-02-25
Published2026-04-01
Last updated2026-09-13 11:58 UTC
CompanyOlympus
DistributionUS

Product

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Reason

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Identifiers

code_infoModel: 3005PK; UDI: 00821925036000; Lot Numbers: All unexpired;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1642-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.