FDA Device recall Z-1635-2026
MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
- FDA Device
- Class II
- ongoing
- Published 2026-04-01
On 2026-04-01 the FDA classified a Class II recall of MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61 by Medtronic Minimed, citing three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which co….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1635-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-01-30 |
| Published | 2026-04-01 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medtronic Minimed |
| Distribution | US |
Product
MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61
Reason
Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).
Identifiers
| code_info | MiniMed 780G Model Number: MMT-1884 UDI-DI code: 763000414344 Serial Numbers: NG2445689H NG2445778H NG2455039H NG246152…MiniMed 780G Model Number: MMT-1884 UDI-DI code: 763000414344 Serial Numbers: NG2445689H NG2445778H NG2455039H NG2461527H NG2461608H NG2471389H NG2471573H NG2472820H NG2473989H NG2480622H NG2481237H NG2488091H NG2489139H NG2489358H NG2493893H NG2494899H NG2496212H NG2496474H NG2497261H NG2499391H NG2499496H NG2500475H NG2500548H NG2501835H NG2502145H NG2502898H NG2507769H NG2507907H NG2508301H NG2508631H NG2508977H NG2511996H NG2513292H NG2521020H NG2522070H NG2523184H NG2530419H NG2531373H NG2531462H NG2531531H NG2532792H NG2533855H NG2535235H NG2535576H NG2537531H NG2538456H NG2540108H NG2541102H NG2541837H NG2543029H NG2544725H NG2548997H NG2550437H NG2570180H NG2570258H NG2570863H NG2576778H NG2579113H NG2579956H NG2580323H NG2580479H NG2583059H NG2583537H NG2586717H NG2589082H NG2590426H NG2603812H NG2603854H NG2604027H NG2611960H NG2612797H NG2614177H NG2617670H NG2617983H NG2629660H NG2635544H NG2636572H NG2637401H NG2642169H NG2644227H NG2644567H NG2647406H NG2647913H NG2648318 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1635-2026%22