RecallRadar

FDA Device recall Z-1632-2023

Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating…

On 2023-05-31 the FDA classified a Class II recall of Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating… by Siemens Healthcare Diagnostics, citing reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1632-2023
ClassificationClass II
Statusongoing
Initiated2023-04-05
Published2023-05-31
Last updated2026-09-13 11:56 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica¿ CH Analyzer Siemens Material Number (SMN): 11099440

Reason

Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

Identifiers

code_infoUDi-DI: 00630414597614 Lot Numbers: 74285306 Exp. Date: 2023-05-31; 74536672 Exp Date: 2024-03-31; 74732386 Exp. Date: 2024-11-30

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1632-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.