FDA Device recall Z-1627-2016

GEM(TM), FLOWCOUPLER(R), 2.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2752-FC Product Usage: The FlowCOUPLER Device is a single use,…

On 2016-05-18 the FDA classified a Class II recall of GEM(TM), FLOWCOUPLER(R), 2.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2752-FC Product Usage: The FlowCOUPLER Device is a single use,… by Baxter Healthcare, citing instructions for use booklet may puncture the outer Tyvek lid.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1627-2016
ClassificationClass II
Statusterminated
Initiated2016-04-21
Published2016-05-18
Last updated2026-09-13 11:54 UTC
CompanyBaxter Healthcare
DistributionUS

Product

GEM(TM), FLOWCOUPLER(R), 2.0 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2752-FC Product Usage: The FlowCOUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FlowCOUPLER Device is used in conjunction with the FlowCOUPLER Monitor, the FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site.

Reason

Instructions for use booklet may puncture the outer Tyvek lid.

Identifiers

code_infoSP15H05-1067843 thru SP15K05 -1094463