RecallRadar

FDA Device recall Z-1623-2015

CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985

On 2015-05-27 the FDA classified a Class II recall of CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985 by Ge Healthcare, citing GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scann….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1623-2015
ClassificationClass II
Statusterminated
Initiated2015-04-22
Published2015-05-27
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare
DistributionFL, MN, TX

Product

CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state detector.

Reason

GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res

Identifiers

code_infoMfg Lot or Serial # System ID 00000331100CN8 904223CT 00000345975CN7 409838LS16 00000337343CN8 512OCCT 00000344962CN6 INSIGHTPCT1356

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1623-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.