Medtronic, REF: MMT-7008D, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 1PK

On 2021-05-26 the FDA classified a Class II recall of Medtronic, REF: MMT-7008D, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 1PK by Medtronic, citing use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1618-2021
ClassificationClass II
Statusongoing
Initiated2021-02-26
Published2021-05-26
Last updated2026-09-13 11:56 UTC
CompanyMedtronic
Distributionn/a

Product

Medtronic, REF: MMT-7008D, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 1PK; GTIN: 00763000252045 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Reason

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Identifiers

code_infoAll Lot numbers within active shelf life