FDA Device recall Z-1616-2020

Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - P…

On 2020-04-08 the FDA classified a Class II recall of Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - P… by Philips North America, citing menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1616-2020
ClassificationClass II
Statusterminated
Initiated2020-02-20
Published2020-04-08
Last updated2026-09-13 11:55 UTC
CompanyPhilips North America
DistributionUS

Product

Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.

Reason

Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

Identifiers

code_infoSoftware version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00