FDA Device recall Z-1606-2016
Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System
- FDA Device
- Class II
- terminated
- Published 2016-05-11
On 2016-05-11 the FDA classified a Class II recall of Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System by Carefusion 303, citing the Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1606-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-03-24 |
| Published | 2016-05-11 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Carefusion 303 |
| Distribution | US |
Product
Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
Reason
The Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.
Identifiers
| code_info | Serial no. 13734283 13734308 13734309 13734310 13734312 13734313 13734337 13734339 13734345 13734351 1373435…Serial no. 13734283 13734308 13734309 13734310 13734312 13734313 13734337 13734339 13734345 13734351 13734352 13734353 13734354 13734355 13734361 13734364 13734365 13734366 13734367 13734368 13734370 13734374 13734376 13734377 13734379 13734382 13734383 13734408 13734413 13734414 13734415 13734418 13734424 13734427 13734430 13734431 13734432 13734435 13734437 13734440 13734441 13734452 13734453 13734454 13734455 13734456 13734457 13734458 13734459 13734460 13734461 13734462 13734463 13734464 13734465 13734466 13734467 13734468 13734469 13734470 13734474 13734476 13734479 13734486 13734487 13734488 13734489 13734491 13734496 13734498 13734510 13734511 13734512 13734518 13734559 13734560 13734582 13734583 13734584 13734585 13734586 13734589 13734590 13734591 13734595 13734600 13734601 13734607 13734608 13734613 13734614 13734621 13734626 13734627 13734630 13734634 13734635 13734636 13734640 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1606-2016%22