RecallRadar

FDA Device recall Z-1605-2019

Maquet Getinge-BO-T 46201 Small Pediatric Cardiac Surge Material: 709000133

On 2019-06-05 the FDA classified a Class II recall of Maquet Getinge-BO-T 46201 Small Pediatric Cardiac Surge Material: 709000133 by Maquet Cardiovascular, citing custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1605-2019
ClassificationClass II
Statusterminated
Initiated2019-04-18
Published2019-06-05
Last updated2026-09-13 11:55 UTC
CompanyMaquet Cardiovascular
DistributionUS

Product

Maquet Getinge-BO-T 46201 Small Pediatric Cardiac Surge Material: 709000133

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Identifiers

code_infoBatch Numbers: 3000044424 3000046488 3000049459 3000055093 3000060357

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.