FDA Device recall Z-1604-2026
I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbe…
- FDA Device
- Class II
- ongoing
- Published 2026-04-01
On 2026-04-01 the FDA classified a Class II recall of I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbe… by I T S, citing updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1604-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-01-23 |
| Published | 2026-04-01 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | I T S |
| Distribution | US |
Product
I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Identifiers
| code_info | 1. Article Number (Lot Numbers): 21232-3 (26/F65636), 21232-4 (26/F61722, 26/F65637), 21232-6 (26/F62564, 26/F65638), 2…1. Article Number (Lot Numbers): 21232-3 (26/F65636), 21232-4 (26/F61722, 26/F65637), 21232-6 (26/F62564, 26/F65638), 21233-3 (26/F65639), 21233-4 (26/F65640), 21233-6 (26/F61723, 26/F65046), 21234-4 (26/F65641, 26/F66504), 21235-4 (26/F65642, 26/F67810). 2. Article Number (Lot Numbers): 21231-6 (643/012301, 643/022310, 643/032308, 643/0630, 643/0657, 643/102402, 643/122301). 3. Article Number (Lot Numbers): 37302-18 (18/636447), 37302-20 (18/802189). |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1604-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.