FDA Device recall Z-1600-2026
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw
- FDA Device
- Class II
- ongoing
- Published 2026-04-01
On 2026-04-01 the FDA classified a Class II recall of Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw by I T S, citing updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1600-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-01-23 |
| Published | 2026-04-01 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | I T S |
| Distribution | US |
Product
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Identifiers
| code_info | 1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736). |
|---|