FDA Device recall Z-1597-2018

Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner

On 2018-05-09 the FDA classified a Class II recall of Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner by Siemens Medical Solutions Usa, citing error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstru….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1597-2018
ClassificationClass II
Statusterminated
Initiated2018-01-26
Published2018-05-09
Last updated2026-09-13 11:54 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.

Reason

Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.

Identifiers

code_infoSoftware version VJ20A. Code No. 10532746, 10532748.