FDA Device recall Z-1596-2016

VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712

On 2016-05-11 the FDA classified a Class II recall of VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712 by Biomerieux, citing product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1596-2016
ClassificationClass II
Statusterminated
Initiated2016-02-24
Published2016-05-11
Last updated2026-09-13 11:54 UTC
CompanyBiomerieux
Distributionn/a

Product

VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Durham, North Carolina 27712. VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria. The VITEK 2 System is not considered a life supporting/sustaining device and does not deliver any type of energy or substance to the patient. The VITEK 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.

Reason

Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.

Identifiers

code_infoLot/Serial: 485368220 Expiry: 12-JAN-17