FDA Device recall Z-1593-2024
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
- FDA Device
- Class II
- ongoing
- Published 2024-04-24
On 2024-04-24 the FDA classified a Class II recall of Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567 by Randox Laboratories, citing microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Qua….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1593-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-01-31 |
| Published | 2024-04-24 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Randox Laboratories |
| Distribution | US |
Product
Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Reason
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Identifiers
| code_info | GTIN: 05055273204520 (1) Batch/Lot: 619282 / 1094MA-1098MA |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.