FDA Device recall Z-1589-2022
Drugs of Abuse Tests: a. Wondfo Tdip Drugs of Abuse Test Tdip Single Drug Fentanyl (FEN) Urine Sample 25 Tests
- FDA Device
- Class II
- ongoing
- Published 2022-08-31
On 2022-08-31 the FDA classified a Class II recall of Drugs of Abuse Tests: a. Wondfo Tdip Drugs of Abuse Test Tdip Single Drug Fentanyl (FEN) Urine Sample 25 Tests by Mckesson Medical Surgical Corporate Office, citing facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1589-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-05-25 |
| Published | 2022-08-31 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Mckesson Medical Surgical Corporate Office |
| Distribution | US |
Product
Drugs of Abuse Tests: a. Wondfo Tdip Drugs of Abuse Test Tdip Single Drug Fentanyl (FEN) Urine Sample 25 Tests. Model Number: AWFTY-114. b. Alere Drugs of Abuse Test E-Z Split Key Cup 12-Drug Panel AMP, BAR, BUP, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PPX, THC Urine Sample 25 Tests Model Number: DOA-1127-041-19. c. Instant Technologies, Inc. Drugs of Abuse Test iCup A.D. 5-Drug Panel with Adulterants AMP, COC, mAMP/MET, OPI, THC (OX, pH, SG) Urine Sample 25 Tests Model Number: I-DUA-157-013. d. Instant Technologies, Inc. Drugs of Abuse Test iCup A.D. 6-Drug Panel with Adulterants AMP, BZO, COC, mAMP/MET, OPI, THC, (OX, pH, SG) Urine Sample Tests Model Number: I-DUA-167-022. e. Alere Drugs of Abuse Test iCup Dx Pro 2 10-Drug Panel with Adulterants AMP, BAR, BUP, BZO, COC, mAMP/MET, OPI300, OXY, MTD, THC (CR, OX, SG) Urine Sample 25 Tests Model Number: I-DXP-2107-01. f. Alere Drugs of Abuse Test iScreen 5-Drug Panel AMP, COC, mAMP/MET, OPI, THC Urine Sample 25 Tests Model Number: IS5M. g. Instant Technologies, Inc. Drugs of Abuse Test T-Cube 7-Drug Panel AMP50, COC20, mAMP/MET50, OPI40, OXY20, PCP10, THC40 Saliva Sample 25 Tests Model Number: TCUBE-7PCPB. h. Wondfo Drugs of Abuse Test with Adulterants Tcup 14-Drug Panel AMP1000, BAR300, BZO300, BUP10, COC300, mAMP/MET1000, MDMA500, MTD300, MOP(OPI300), OXY100, PCP25, PPX300, TCA1000, THC50 (CR, pH, SG) Urine Sample 25 Tests Model Number: TDOA-1145A3. i. Instant Technologies, Inc. Drugs of Abuse Test UScreen 12-Drug Panel with Adulterants AMP, BAR, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PCP, TCA, THC, (CR, pH, SG) Urine Sample 25 Tests Model Number: USSCUPA-12CLIA. j. Instant Technologies, Inc. Drugs of Abuse Test UScreen 12-Drug Panel with Adulterants AMP, BAR, BUP, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PCP, THC (CR, pH, SG) Urine Sample 25 Tests Model Number: USSCupA-12BUP300. k. Alere Drugs of Abuse Test E-Z Split Key Cup 6-Drug Panel AMP, BZO, COC, mAMP/MET, OPI, THC Urine Sample 25 Tests Model Number: DOA-2167-019. l. Alere Drugs of Abuse Test E-Z Split Key Cup A.D. 5-Drug Panel with Adulterants AMP, COC, mAMP/MET, OPI, THC (CR, GL, NI, OX, pH, SG) Urine Sample 25 Tests Model Number: DUD-157-012-019. m. WONDFO USA CO., LTD. Wondfo Drug Test Dip Card Drugs of Abuse Test Tdip Single Drug Buprenorphine (BUP) Urine Sample 25 Tests Model Number: WDBU-114.
Reason
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Identifiers
| code_info | a. GTIN: Unknown All lots received between 7/4/21-9/30/21. b. GTIN:00815845020782; 10815845020789 All lots receiv…a. GTIN: Unknown All lots received between 7/4/21-9/30/21. b. GTIN:00815845020782; 10815845020789 All lots received between 7/4/21-9/30/21. c. GTIN:00858945006072; 10858945006079 All lots received between 7/4/21-9/30/21. d. GTIN: 00858945006133; 10858945006130 All lots received between 7/4/21-9/30/21. e. GTIN: 00817405020355; 00858945006829 All lots received between 7/4/21-9/30/21. f. GTIN: 00858945006508; 10858945006505 All lots received between 7/4/21-9/30/21. g. GTIN: 00817405020959 All lots received between 7/4/21-9/30/21. h. GTIN: 00816862020076; 10816862020073; 00816862029079; 00816862025071 All lots received between 7/4/21-9/30/21. i. GTIN: 00817405021192 All lots received between 7/4/21-9/30/21. j. GTIN: 00858998006104 All lots received between 7/4/21-9/30/21. k. GTIN: 00815845020720; 10815845020727 All lots received between 7/4/21-9/30/21. l. GTIN: 00815845020355; 10815845020352. m. GTIN: 00816862025415. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1589-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.