FDA Device recall Z-1583-2022
a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD. Part No. 0…
- FDA Device
- Class II
- ongoing
- Published 2022-08-31
On 2022-08-31 the FDA classified a Class II recall of a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD. Part No. 0… by Mckesson Medical Surgical Corporate Office, citing facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1583-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-05-25 |
| Published | 2022-08-31 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Mckesson Medical Surgical Corporate Office |
| Distribution | US |
Product
a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD. Part No. 0356-50. b. Coagulation Control Kit Coag-Sense PT / INR Test (Prothrombin Time Test) Low Level / High Level 10 Strips Part No. 03P69-10.
Reason
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Identifiers
| code_info | a. GTIN: 00861900000336; 10861900000395; 00861900000398. Lot codes distributed between 5/30/2021 - 8/24/2021. b.…a. GTIN: 00861900000336; 10861900000395; 00861900000398. Lot codes distributed between 5/30/2021 - 8/24/2021. b. GTIN 00861900000343 Lots Distributed: 8/20/2021 - 8/20/2021 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1583-2022%22