FDA Device recall Z-1582-2023
DuraLife Autoclavable Silicone Extension Tube, REF 60-1510 (UltraSet Product Code 66-2505)
- FDA Device
- Class II
- ongoing
- Published 2023-05-24
On 2023-05-24 the FDA classified a Class II recall of DuraLife Autoclavable Silicone Extension Tube, REF 60-1510 (UltraSet Product Code 66-2505) by Smiths Medical Asd, citing duraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning m….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1582-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-04-05 |
| Published | 2023-05-24 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Smiths Medical Asd |
| Distribution | US |
Product
DuraLife Autoclavable Silicone Extension Tube, REF 60-1510 (UltraSet Product Code 66-2505); Swivel Elbow with Suction Port for use with breathing circuits
Reason
DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.
Identifiers
| code_info | Lot Numbers: 3572938, 3709023, 3917565, 4081673, 4176083, 4258736, 3582668, 3735447, 3936716, 4090325, 4176084, 427270…Lot Numbers: 3572938, 3709023, 3917565, 4081673, 4176083, 4258736, 3582668, 3735447, 3936716, 4090325, 4176084, 4272705, 3602084, 3784974, 3974795, 4100952, 4187591, 4272706, 3617047, 3812333, 4011957, 4104648, 4207974, 4284609, 3623883, 3828394, 4029611, 4119019, 4207975, 4290273, 3631678, 3836006, 4038969, 4138936, 4220676, 4333910, 3662331, 3874536, 4063254, 4148601, 4242511, 4340649, 3688907, 3884430, 4067164, 4160298, 4254341, 4348704, 3693346, 3904854, 4076803, 4167797, 4258735 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1582-2023%22