Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

On 2019-06-05 the FDA classified a Class II recall of Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551 by Maquet Cardiovascular, citing custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1579-2019
ClassificationClass II
Statusterminated
Initiated2019-04-18
Published2019-06-05
Last updated2026-09-13 11:55 UTC
CompanyMaquet Cardiovascular
DistributionUS

Product

Maquet Getinge-T 41303 AV MANIFOLD FOR ECC PA Material:701066551

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Identifiers

code_infoBatch Numbers: 3000041542 3000046483 3000049065 3000055328 3000057682 3000060483 3000063712

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.